Remote Commands
R&S
®
FPL1000
720
User Manual 1178.3370.02 ─ 09
CALC:MARK:X:SLIM ON
//Activates limit lines for evaluation.
CALC:MARK:X:SLIM:LEFT 1ms
//Sets the left limit line to 326 µs.
CALC:MARK:X:SLIM:RIGH 6ms
//Sets the right limit line to 538 µs.
//--------------Performing the Measurement------------------------
INIT;*WAI
//Initiates the measurement and waits until the measurement is finished.
//-------------Retrieving the Results--------------------------------
CALC:MARK:FUNC:SUMM:MEAN:RES?
CALC:MARK:FUNC:SUMM:PPE:RES?
CALC:MARK:FUNC:SUMM:RMS:RES?
//Queries the mean, peak and RMS time domain power.
11.6.11
Measuring the Harmonic Distortion
All remote control commands specific to harmonic distortion measurements are descri-
bed here.
..................................................................................720
...............................................................................721
................................................................................722
.................................................................................................722
Example: Measuring the Harmonic Distortion
.......................................................723
11.6.11.1
Activating the Measurement
The following command activates harmonic distortion measurement.
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics[:STATe]
.............................................. 720
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics[:STATe]
<State>
This command turns the harmonic distortion measurement on and off.
Note the following:
●
If you perform the measurement in the frequency domain, the search range for the
frequency of the first harmonic, whose power is determined, is defined by the last
span.
●
If you perform the measurement in the time domain, the current center frequency is
used as the frequency of the first harmonic. Thus, the frequency search is
bypassed. The first harmonic frequency is set by a specific center frequency in
zero span before the harmonic measurement is started.
Configuring and Performing Measurements