Remote Commands
R&S
®
FPL1000
740
User Manual 1178.3370.02 ─ 09
<RFAttenuation>
Defines the RF attenuation for a list entry.
Range:
0 to 70
Increment: 1
Default unit: dB
<FilterType>
Selects the filter type for a list entry. For more information see
[SENSe:]BANDwidth[:RESolution]:TYPE
<RBW>
Defines the resolution bandwidth for a list entry.
<VBW>
Defines the measurement time for a list entry.
<MeasTime>
Defines the measurement time for a list entry.
Range:
1 µs to 16000 s
Default unit: s
<TriggerLevel>
The trigger level must be 0.
<PowerLevel>
Power level for each frequency included in the measurement.
The command returns up to 3 power levels for each frequency,
depending on the number of evaluation modes you have turned
on with
The result is a list of floating point values separated by commas.
The unit depends on
[SENSe:]LIST:POWer:SET
<State>, <State>, <State>, <TriggerSource>,
<TriggerSlope>, <TriggerOffset>, <GateLength>
This command defines global List Evaluation parameters.
These parameters are valid for every frequency you want to measure.
The state of the first three parameters (<PeakPower>, <RMSPower> and <AVG-
Power>) define the number of results for each frequency in the list.
Note that you have to set the trigger level
after
sending this command.
Parameters:
<State>
ON | OFF | 0 | 1
Turns peak power evaluation on and off.
*RST:
1
<State>
ON | OFF | 0 | 1
Turns RMS power evaluation on and off.
*RST:
0
<State>
ON | OFF | 0 | 1
Turns average power evaluation on and off.
*RST:
0
<TriggerSource>
IMMediate | LINE | EXTernal | VIDeo | RFPower | IFPower |
RFPower | EXT2 | EXT3 | EXT4 | LXI
Selects a trigger source.
Configuring and Performing Measurements