![Ronde & Schwarz 1304.0004.03 User Manual Download Page 670](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985670.webp)
Remote Commands
R&S
®
FPL1000
670
User Manual 1178.3370.02 ─ 09
Parameters:
<Level>
Relative limit at the start frequency of a SEM range.
Range:
-400 to 400
*RST:
-50
Default unit: dBc
Example:
SENS:ESP:RANG:LIM:REL:STAR -10
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt:ABS
<Level>
This command defines an absolute limit for the MAX function of the relative limit for a
SEM range.
For more information see
"Relative limit line functions"
Suffix:
<sb>
.
1 to 3
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Level>
Absolute limit at the start frequency of a SEM range to be used
in addition to the relative limit if the MAX function is enabled (see
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:
RELative:STARt:FUNCtion
Range:
-400 to 400
*RST:
-13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STARt:
ABSolute -10
Chapter 11.6.7.10, "Example: SEM
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt:FUNCtion
<Function>
Suffix:
<sb>
.
1 to 3
<ri>
1..n
Selects the SEM range.
Configuring and Performing Measurements