![Ronde & Schwarz 1304.0004.03 User Manual Download Page 177](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985177.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
177
User Manual 1178.3370.02 ─ 09
Preamp
Switches the preamplifier on or off.
For details on the preamplifier, see
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:GAIN:STATe
Transducer Factor
Sets a transducer for the specified range. You can only choose a transducer that fulfills
the following conditions:
●
The transducer overlaps or equals the span of the range.
●
The x-axis is linear.
●
The unit is dB.
For details on transducers, see
Chapter 9.7.1, "Basics on Transducer Factors"
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:TRANsducer
Limit Check <n>
Sets the type of limit check for the n-th power class in the range. Up to four limits are
possible.
For details on limit checks, see
"Limit Lines in SEM Measurements"
The limit state affects the availability of all limit settings.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:STATe
Abs Limit Start / Stop <n>
Sets an absolute limit value for the n-th power class at the start or stop frequency of
the range [dBm].
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:ABSolute:STARt
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:ABSolute:STOP
Rel Limit Start / Stop <n>
Sets a relative limit value for the n-th power class at the start or stop frequency of the
range [dBc].
By default, this value is a fixed relative limit, i.e. no function is defined. To define a
function for the relative limit, select the input field for "Rel Limit Start" or "Rel Limit
Stop" and then the "f(x)" icon that appears.
Measurements and Results