General Instrument Setup
R&S
®
FPL1000
517
User Manual 1178.3370.02 ─ 09
....................................................................................... 517
Out-of-range value behavior
By default, if you enter a value that is outside the valid range in an input field for a set-
ting, a warning is displayed and the value is not accepted. Alternatively, entries below
the minimum value can automatically be set to the minimum possible entry, and entries
above the maximum value set to the maximum possible entry. This behavior avoids
errors and facilitates setting correct values.
9.4.6
Additional Interfaces
The following settings are only available if the "Additional interfaces" option R&S
FPL1-
B5 is installed.
Wait for Trigger Polarity
The optional AUX PORT connector of the R&S
FPL1000 can provide a signal that indi-
cates the instrument is ready to receive a trigger signal.
(For details on the connector see the R&S
FPL1000 Getting Started manual).
The signal polarity that indicates the trigger availability is configurable.
"Low"
A low signal (=
0
V) indicates the instrument is ready to receive a trig-
ger.
"High"
A high signal (=
5
V) indicates the instrument is ready to receive a
trigger.
Remote command:
OUTPut:UPORt:WTRigger:POLarity LOW | HIGH
(see the R&S
FPL1000 user
manual)
9.5
Service Functions
Access
: [Setup] > "Service"
Service Functions