Remote Commands
R&S
®
FPL1000
671
User Manual 1178.3370.02 ─ 09
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Function>
OFF | MAX
Defines the function to be used to determine the relative limit
line start value
MAX
The maximum of the relative
and
the absolute level is used as
the limit start value. Use the
RANGe<ri>:LIMit<li>:RELative:STARt
]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:
STARt:ABS
commands to define these values.
OFF
No function is used, the relative limit line is defined by a fixed
relative start value. Use the
RANGe<ri>:LIMit<li>:RELative:STARt
define this value.
*RST:
OFF
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STARt:
FUNCtion MAX
Chapter 11.6.7.10, "Example: SEM
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP
<Level>
This command defines a relative limit for a SEM range.
Unlike manual operation, you can define a relative limit anytime and regardless of the
limit check mode.
Suffix:
<sb>
.
1 to 3
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Level>
Relative limit at the stop frequency of a SEM range.
Range:
-400 to 400
*RST:
-50
Default unit: dBc
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP -15
Chapter 11.6.7.10, "Example: SEM
Configuring and Performing Measurements