The Spectrum Application (RF Measurements)
R&S
®
FPL1000
145
User Manual 1178.3370.02 ─ 09
A programming example demonstrating an ACLR measurement in a remote environ-
ment is provided in
Chapter 11.6.4.8, "Programming Examples for Channel Power
Measurement Example 2 – Measuring Adjacent Channel Power of a W-CDMA
Uplink Signal
Measurement Example 3 – Measuring the Intrinsic Noise of the R&S
.................................................................................146
Measurement Example 2 – Measuring Adjacent Channel Power of a W-CDMA
Uplink Signal
Test setup:
Signal generator settings (e.g. R&S SMW):
Frequency:
1950 MHz
Level:
4 dBm
Modulation:
3GPP W-CDMA Reverse Link
Procedure:
1. Preset the R&S
FPL1000.
2. Set the center frequency to
1950
MHz
.
3. Select the "Channel Power ACLR" measurement function from the "Select Mea-
surement" dialog box.
4. Set the "W-CDMA 3GPP REV" standard for adjacent channel power measurement
in the "ACLR Setup" dialog box.
The R&S
FPL1000 sets the channel configuration to the W-CDMA standard for
mobiles with two adjacent channels above and below the transmit channel. The
frequency span, the resolution and video bandwidth and the detector are automati-
cally set to the correct values. The spectrum is displayed in the upper window and
the channel power, the level ratios of the adjacent channel powers and the channel
configuration in the lower window. The individual channels are displayed as bars in
the graph.
5. Set the optimal reference level and RF attenuation for the applied signal level using
the "Auto Level" function.
Measurements and Results