Remote Commands
R&S
®
FPL1000
661
User Manual 1178.3370.02 ─ 09
Manual operation:
See
11.6.7.2
Controlling the Measurement
The following commands control the measurement itself.
INITiate<n>:ESPectrum
This command initiates a Spectrum Emission Mask measurement.
Suffix:
<n>
.
irrelevant
[SENSe:]SWEep:MODE
<Mode>
This command selects the spurious emission and spectrum emission mask measure-
ments.
You can select other measurements with
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>[:STATe]
Parameters:
<Mode>
LIST | AUTO | ESPectrum
AUTO
Turns on basic spectrum measurements.
ESPectrum
Turns on spectrum emission mask measurements.
LIST
Turns on spurious emission measurements.
*RST:
AUTO
Example:
SWE:MODE ESP
11.6.7.3
Configuring a Multi-SEM Measurement
In the Spectrum application only, spectrum emissions can be measured for multiple
sub blocks of channels (see
"SEM with Multiple Sub Blocks ("Multi-SEM")"
on page 169). Up to 8 sub blocks (with 7 gaps) can be defined. For each sub block, the
familiar configuration settings concerning ranges, limit lines etc. can be defined in indi-
vidual tabs (select the sub block using the <sb> suffix in the corresponding com-
mands). In addition, settings on the sub blocks themselves must be configured.
Useful commands for multi-SEM measurements described elsewhere:
●
[SENSe:]ESPectrum<sb>:RANGe<ri>:MLCalc
Configuring and Performing Measurements