Remote Commands
R&S
®
FPL1000
630
User Manual 1178.3370.02 ─ 09
11.6.2
Configuring Power Measurements
The following commands work for several power measurements.
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:MODE
............................................... 630
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
.............................................630
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:SELect
.............................................. 632
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>[:STATe]
..............................................632
................................................................................ 633
[SENSe:]POWer:ACHannel:PRESet:RLEVel
.....................................................................633
.................................................................................................633
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:MODE
<Mode>
This command selects the trace display mode for power measurements.
Suffix:
<n>
.
<m>
<sb>
irrelevant
Parameters:
<Mode>
WRITe | MAXHold
WRITe
The power is calculated from the current trace.
MAXHold
The power is calculated from the current trace and compared
with the previous power value using a maximum algorithm.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
<Measurement>
This command queries the results of power measurements.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Configuring and Performing Measurements