![Ronde & Schwarz 1304.0004.03 User Manual Download Page 648](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985648.webp)
Remote Commands
R&S
®
FPL1000
648
User Manual 1178.3370.02 ─ 09
CALCulate<n>:LIMit<li>:ACPower:ACHannel:ABSolute:STATe
This command turns the absolute limit check for the adjacent channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Parameters:
<State>
Manual operation:
See
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]
<LowerLimit>[,
<UpperLimit>]
This command defines the relative limit of the adjacent channels. The reference value
for the relative limit is the measured channel power.
If you have defined an absolute limit as well as a relative limit, the R&S
FPL1000 uses
the lower value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Parameters:
<LowerLimit>
The limit of the lower adjacent channel.
Range:
0 dB to 100 dB
*RST:
0 dB
Default unit: dB
<UpperLimit>
The limit of the upper adjacent channel.
Range:
0 dB to 100 dB
*RST:
0 dB
Default unit: dB
Manual operation:
See
CALCulate<n>:LIMit<li>:ACPower:ACHannel:RESult?
This command queries the state of the limit check for the adjacent channels in an
ACLR measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Configuring and Performing Measurements