The Spectrum Application (RF Measurements)
R&S
®
FPL1000
253
User Manual 1178.3370.02 ─ 09
Initial Measurement - Peak Search
.......................................................................259
Final Measurement at the Marker Position
........................................................... 260
Resolution Bandwidth and Filter Types
EMI testing requires resolution filters with a 6
dB bandwidth. The R&S
FPL1000 EMI
measurement adds the following bandwidths that comply to commercial and military
standards to those already available with the base unit:
Commercial (CISPR, FFC etc.)
●
200
Hz
●
9
kHz
●
120
kHz
●
1
MHz (not with Quasipeak detector, see
"Quasipeak detector (CISPR filter only)"
Military (MIL Std)
●
10
Hz
●
100
Hz
●
1
kHz
●
10
kHz
●
100
kHz
●
1
MHz
For the Quasipeak, CISPR Average, or RMS Average detector, the bandwidth is fixed
depending on the frequency. For more information see
Detectors and Dwell Time
The EMI measurement adds new detectors to those already available with the base
unit. The additional detectors are especially designed for and required by EMI applica-
tions.
The additional detectors are available only if the EMI (R&S
FPL1-K54) measurement
option is installed and the filter type "CISPR" or "MIL" is selected (see
on page 253). The additional detectors can be selected for
the final test in the EMI measurement, or as trace detectors in zero span measure-
ments. For the initial peak search in the EMI measurement, these detectors are
not
available.
The detector to be used for the initial peak search is configured in the trace settings
(see
Chapter 7.10.1.2, "Trace Settings"
on page 393), while the detector for the final
test is configured in the EMI marker settings, see
Measurements and Results