Remote Commands
R&S
®
FPL1000
580
User Manual 1178.3370.02 ─ 03
Return values:
<RefRange>
Number of the current reference range.
Range:
1 to 30
Usage:
Query only
[SENSe:]ESPectrum<sb>:RTYPe
<Type>
This command defines the type of the power reference.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
Parameters:
<Type>
PEAK
Measures the highest peak within the reference range.
CPOWer
Measures the channel power within the reference range (integral
bandwidth method).
*RST:
CPOWer
Manual operation:
See
8.6.6.6
Configuring the Power Classes
The following commands define the power classes for SEM measurements.
CALCulate<n>:LIMit<k>:ESPectrum<sb>:LIMits
................................................................580
CALCulate<n>:LIMit<k>:ESPectrum<sb>:MODE
...............................................................581
CALCulate<n>:LIMit<k>:ESPectrum<sb>:VALue
...............................................................582
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass:COUNt
..................................................582
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass<PClass>[:EXCLusive]
............................ 583
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass<PClass>:LIMit[:STATe]
.......................... 583
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass<PClass>:MAXimum
............................... 584
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass<PClass>:MINimum
................................ 584
CALCulate<n>:LIMit<k>:ESPectrum<sb>:LIMits
<Max1>,<Max2>,<Max3>
This command sets or queries up to 4 power classes in one step. You can only define
values for the number of power classes defined by
Suffix:
<n>, <k>
.
irrelevant
<sb>
1 | 2 | 3
Configuring and Performing Measurements