Remote Commands
R&S
®
FPL1000
665
User Manual 1178.3370.02 ─ 03
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST:
-10 dBm
Example:
TRIG:LEV:IFP -30DBM
Manual operation:
See
TRIGger[:SEQuence]:LEVel:IQPower
<TriggerLevel>
This command defines the magnitude the I/Q data must exceed to cause a trigger
event. Note that any RF attenuation or preamplification is considered when the trigger
level is analyzed.
Parameters:
<TriggerLevel>
Range:
-130 dBm to 30 dBm
*RST:
-20 dBm
Example:
TRIG:LEV:IQP -30DBM
Manual operation:
See
TRIGger[:SEQuence]:SLOPe
<Type>
For all trigger sources except time you can define whether triggering occurs when the
signal rises to the trigger level or falls down to it.
Parameters:
<Type>
POSitive | NEGative
POSitive
Triggers when the signal rises to the trigger level (rising edge).
NEGative
Triggers when the signal drops to the trigger level (falling edge).
*RST:
POSitive
Example:
TRIG:SLOP NEG
Manual operation:
See
TRIGger[:SEQuence]:SOURce
<Source>
This command selects the trigger source.
For details on trigger sources see
Note on external triggers:
If a measurement is configured to wait for an external trigger signal in a remote control
program, remote control is blocked until the trigger is received and the program can
continue. Make sure this situation is avoided in your remote control programs.
Setting Basic Spectrum RF Measurement Parameters