The Spectrum Application (RF Measurements)
R&S
®
FPL1000
399
User Manual 1178.3370.02 ─ 03
"Pulse[0.1]"
"Video Bandwidth" = 10 x "Resolution Bandwidth"
or
"Video Bandwidth" = "10 MHz" (= max. VBW)
Recommended for pulse signals
"Noise[10]"
"Video Bandwidth" = "Resolution Bandwidth/10"
Recommended for noise measurements
"Manual"
The coupling ratio is defined manually.
The RBW/VBW ratio can be set in the range of 0.001 to 1000.
Remote command:
Filter Type
Defines the filter type.
The following filter types are available:
●
Normal (3dB)
●
Channel
For more information see
Chapter 7.6.1.6, "Which Data May Pass: Filter Types"
Remote command:
[SENSe:]BANDwidth[:RESolution]:TYPE
Default Coupling
Sets all coupled functions to the default state ( "Auto" ). In addition, the ratio "RBW/
VBW" is set to "Sine[1/1]" and the ratio "Span/RBW" to
100.
For more information see
Chapter 7.6.1.3, "Coupling VBW and RBW"
Remote command:
[SENSe:]BANDwidth[:RESolution]:AUTO
Sweep/Average Count
Defines the number of sweeps to be performed in the single sweep mode. Values from
0 to 200000 are allowed. If the values 0 or 1 are set, one sweep is performed.
The sweep count is applied to all the traces in all diagrams.
If the trace modes "Average" , "Max Hold" or "Min Hold" are set, this value also deter-
mines the number of averaging or maximum search procedures.
In continuous sweep mode, if "Sweep Count" = 0 (default), averaging is performed
over 10 sweeps. For "Sweep Count" =1, no averaging, maxhold or minhold operations
are performed.
For more information, see
Chapter 7.6.1.8, "How Much Data is Measured: Sweep
Bandwidth, Filter and Sweep Configuration