Parameters:
<State>
ON | OFF | 1 | 0
Relative limit check for lower adjacent channel
*RST:
0
<State>
ON | OFF | 1 | 0
Relative limit check for upper adjacent channel
*RST:
0
Manual operation:
See
CALCulate<n>:LIMit<k>:ACPower:ALTernate<ch>:ABSolute
<LowerLimit>,
<UpperLimit>
This command defines the absolute limit of the alternate channels.
If you have defined an absolute limit as well as a relative limit, the R&S
FPL1000 uses
the lower value for the limit check.
Suffix:
<n>, <k>
.
irrelevant
<ch>
1 to 11
Alternate channel number
Parameters:
<LowerLimit>
The limit of the lower adjacent channel.
Range:
-200 dBm to 200 dBm
*RST:
-200 dBm
<UpperLimit>
The limit of the upper adjacent channel.
Range:
-200 dBm to 200 dBm
*RST:
-200 dBm
Manual operation:
See
CALCulate<n>:LIMit<k>:ACPower:ALTernate<ch>:ABSolute:STATe
This command turns the absolute limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<k>:ACPower[:STATe]
.
Suffix:
<n>, <k>
.
irrelevant
<ch>
1 to 11
Alternate channel number
Manual operation:
See
Configuring and Performing Measurements