The Spectrum Application (RF Measurements)
R&S
®
FPL1000
319
User Manual 1178.3370.02 ─ 03
5. In the file selection dialog box, select a storage location and filename for the result
file.
6. Select the "Save" button.
The file with the specified name and the extension
.dat
is stored in the defined
storage location.
7.2.6.7
Measurement Example: Multi-SEM Measurement
The following measurement example demonstrates an SEM measurement for a signal
with multiple sub blocks.
A programming example demonstrating a SEM measurement in a remote environment
is provided in
Chapter 8.6.6.10, "Example: SEM Measurement"
Test setup:
Signal generator settings (e.g. R&S
FPL1000 SMW):
Device
Standard
Center frequency
Level
Test model
SigGen 1
3GPP/FDD
900 MHz
0 dBm
1-16
SigGen 2
EUTRA/LTE
906.5 MHz
0 dBm
1_1_5MHz
Setting up the measurement
1. Preset the R&S
FPL1000.
2. Set the center frequency to
903.25 MHz
.
3. Set the reference level to
10 dBm
with an offset of to
30 dB
.
4. Press the MEAS key or select "Select Measurement" in the "Overview" .
5. Select the "SEM" measurement function.
Measurements and Results