Getting Started
R&S
®
FPL1000
45
User Manual 1178.3370.02 ─ 03
3. Tap the "S Support" softkey.
4. Tap the "Calibration Signal" tab.
5. Tap the "Calibration Frequency RF" option. Leave the frequency at the default
50
MHz.
6. Close the dialog box.
The calibration signal is now sent to the RF input of the R&S
FPL1000. By default,
a continuous frequency sweep is performed, so that the spectrum of the calibration
signal is now displayed in the standard level versus frequency diagram.
Figure 4-4: Calibration signal as RF input
Instrument warmup time
Note that the instrument requires an initial warmup time after switching it on. A mes-
sage in the status bar ("Instrument warming up...") indicates that the operating temper-
ature has not yet been reached. Wait until this message is no longer displayed before
you start a measurement to ensure accurate results.
To optimize the display
To optimize the display for the calibration signal, we will adjust the main measurement
settings.
1. Set the center frequency to the calibration frequency:
a) Tap the "Overview" softkey to display the configuration "Overview" .
b) Tap the "Frequency" button.
c) In the "Center" field, enter
50
on the number pad on the front panel.
d) Press the "MHz" key next to the number pad.
Trying Out the Instrument