Getting Started
R&S
®
FPL1000
46
User Manual 1178.3370.02 ─ 03
2. Reduce the span to 20
MHz:
a) In the "Span" field of the "Frequency" dialog box, enter
20
MHz
.
b) Close the "Frequency" dialog box.
3. Average the trace to eliminate noise:
a) In the configuration "Overview" , tap the "Analysis" button.
b) In the "Traces" tab, select the trace mode "Average" .
c) Enter the "Average Count" :
100
.
d) Close the "Analysis" dialog box.
The display of the calibration signal is now improved. The maximum at the center
frequency (=calibration frequency) of 50
MHz becomes visible.
Figure 4-5: Calibration signal with optimized display settings
4.3.2 Displaying a Spectrogram
In addition to the standard "level versus frequency" spectrum display, the
R&S
FPL1000 also provides a spectrogram display of the measured data. A spectro-
gram shows how the spectral density of a signal varies over time. The x-axis shows the
frequency, the y-axis shows the time. A third dimension, the power level, is indicated
by different colors. Thus you can see how the strength of the signal varies over time for
different frequencies.
1. Tap the "Overview" softkey to display the general configuration dialog box.
2. Tap the "Display Config" button.
The SmartGrid mode is activated, and the evaluation bar with the available evalua-
tion methods is displayed.
3.
Drag the "Spectrogram" icon from the evaluation bar to the diagram area. The blue
area indicates that the new diagram would replace the previous spectrum display.
Since we do not want to replace the spectrum, drag the icon to the lower half of the
display to add an additional window instead.
Trying Out the Instrument