![Rohde & Schwarz FPL1000 Series User Manual Download Page 516](http://html.mh-extra.com/html/rohde-and-schwarz/fpl1000-series/fpl1000-series_user-manual_1477826516.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
516
User Manual 1178.3370.02 ─ 03
Table 7-25: ASCII file format for trace export in the Spectrum application
File contents
Description
Header data
Type;R&S
FPL1000;
Instrument model
Version;1.00;
Firmware version
Date;01.Oct 2006;
Date of data set storage
Mode;ANALYZER;
Operating mode
Preamplifier;OFF
Preamplifier status
Transducer; OFF
Transducer status
Center Freq;55000;Hz
Center frequency
Freq Offset;0;Hz
Frequency offset
Start;10000;Hz
Stop;100000;Hz
Start/stop of the display range.
Unit: Hz for span > 0, s for span = 0, dBm/dB for statistics mea-
surements
Span;90000;Hz
Frequency range (0 Hz in zero span and statistics measure-
ments)
Ref Level;-30;dBm
Reference level
Level Offset;0;dB
Level offset
Rf Att;20;dB
Input attenuation
El Att;2.0;dB
Electrical attenuation
RBW;100000;Hz
Resolution bandwidth
VBW;30000;Hz
Video bandwidth
SWT;0.005;s
Sweep time
Sweep Count;20;
Number of sweeps set
Ref Position;75;%
Position of reference level referred to diagram limits (0 % =
lower edge)
Level Range;100;dB
Display range in y direction. Unit: dB with x-axis LOG, % with x-
axis LIN
x-Axis;LIN;
Scaling of x-axis linear (LIN) or logarithmic (LOG)
y-Axis;LOG;
Scaling of y-axis linear (LIN) or logarithmic (LOG)
x-Unit;Hz;
Unit of x values: Hz with span > 0; s with span = 0; dBm/dB
with statistics measurements
y-Unit;dBm;
Unit of y values: dB*/V/A/W depending on the selected unit
with y-axis LOG or % with y-axis LIN
Data section for individual window
Window;1;Frequency Sweep
Window number and name
Trace 1;;
Selected trace
Importing and Exporting Measurement Results for Evaluation