Remote Commands
R&S
®
FPL1000
581
User Manual 1178.3370.02 ─ 03
Setting parameters:
<Max1>
Defines the value range for power class 1 as
-200 to
<Max1>
.
Only available for
CALC:LIM:ESP:PCL:COUNT >=2
If only 2 power classes are defined, the value range for power
class 2 is defined as
<Max1> to 200
.
Range:
-199 to + 199
<Max2>
Defines the value range for power class 2 as
<Max1> to
<Max2>
.
Only available for
CALC:LIM:ESP:PCL:COUNT >=3
If only 3 power classes are defined, the value range for power
class 3 is defined as
<Max2> to 200
.
Range:
-199 to + 199, <Max2> must be higher than
<Max1>
<Max3>
Defines the value range for power class 3 as
<Max2> to
<Max3>
.
The value range for power class 4 is defined as
<Max3> to
200
.
Only available for
CALC:LIM:ESP:PCL:COUNT = 4
Range:
-199 to + 199, <Max3> must be higher than
<Max2>
Return values:
<RangeLimits>
As a result of the query, the range limits including the beginning
of the first power class (-200) and the end of the last power class
(+200) are indicated.
Example:
CALC:LIM:ESP:LIM -50,50,70
Defines the following power classes:
<-200, -50>
<-50, 50>
<50, 70>
<70, 200>
Query:
CALC:LIM:ESP:LIM?
Response:
-200,-50,50,70,200
CALCulate<n>:LIMit<k>:ESPectrum<sb>:MODE
<Mode>
Which limit line is to be used for an SEM measurement depends on the power class
the input signal power belongs to. This command defines wether the power class is
determined automatically or manually.
Suffix:
<n>, <k>
.
irrelevant
<sb>
1 | 2 | 3
Configuring and Performing Measurements