Getting Started
R&S
®
FPL1000
17
User Manual 1178.3370.02 ─ 03
For details on the connector, refer to
Chapter 4.2.2.2, "AC Power Supply Connection
► Connect the R&S
FPL1000 to the AC power supply using the supplied power
cable.
Since the instrument is assembled in line with the specifications for safety class
EN61010, it must only be connected to an outlet that has a ground contact.
4.1.1.5
Changing the AC Supply Fuse
If the battery option R&S
FPL1-B31 is installed, an AC supply fuse is built into the AC
power supply connector. Only use fuses of the type 5A T IEC60127-2/V (order no.
0099.6735.00). Such fuses are used for all of the specified nominal AC supply vol-
tages.
To change the AC supply fuses
1. Disconnect the power cable.
2. Open the flap covering the fuse holder using a small screwdriver (or similar).
3. Remove the fuse holder.
4. Remove the fuse and install the new one.
5. Reinsert the fuse holder.
6. Close the flap.
4.1.1.6
Connecting an Optional DC Power Supply (R&S
FPL1-B30)
The R&S
FPL1000 can also be equipped with an optional (internal) DC power supply
(R&S
FPL1-B30). If installed, the R&S
FPL1000 can be operated by a DC voltage of
10.4
V to 28
V.
Risk of electric shock
Insufficient insulation and exceeding the current limitation of external power supplies
for safety extra-low DC voltage (SELV) can lead to electric shock.
Be sure to meet the requirements for reinforced/double insulation in accordance with
DIN/EN/IEC 61010 (UL 3111, CSA C22.2 No. 1010.1) orDIN/EN/IEC 60950 (UL 1950,
CSA C22.2 No. 950). Provide current limitation in accordance with DIN EN 61010-1
Appendix F2.1.
Preparing for Use