Remote Commands
R&S
®
FPL1000
821
User Manual 1178.3370.02 ─ 03
Before you can use the command, you have to select and turn on a transducer.
Parameters:
<Comment>
*RST:
(empty comment)
Manual operation:
See
[SENSe:]CORRection:TRANsducer:DATA
<Frequency>,<Level>[,<Frequency>,<Level>]
This command configures transducer factors for specific trace points. A set of trans-
ducer factors defines an interpolated transducer line and can be stored on the instru-
ment.
Parameters:
<Frequency>
The unit for <Frequency> is Hz, which may or may not be omit-
ted. Frequencies have to be sorted in ascending order.
<Level>
The unit for <Level> depends on
Example:
SENSe1:CORRection:TRANsducer:UNIT 'DB'
// Frequency Span 0 Hz to 4 Ghz
SENSe1:CORRection:TRANsducer:DATA 0,8,2e6,5,4GHz,3
Creates the transducer points:
Manual operation:
See
Frequency
Level
0 Hz
8 dB
2 GHz
5 dB
4 GHz
3 dB
[SENSe:]CORRection:TRANsducer:DELete
This command deletes the currently selected transducer factor.
Before you can use the command, you have to select a transducer.
Example:
CORR:TRAN:DEL
Usage:
Event
Manual operation:
See
[SENSe:]CORRection:TRANsducer:SCALing
<ScalingType>
This command selects the frequency scaling of the transducer factor.
Parameters:
<ScalingType>
LINear | LOGarithmic
*RST:
LINear
Configuring the R&S
FPL1000