8.8.8.3
Configuring and Performing a Marker Search
The following commands control the marker search.
CALCulate<n>:MARKer<m>:LOEXclude
.......................................................................... 716
CALCulate<n>:MARKer<m>:PEXCursion
.........................................................................716
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
................................................................717
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
....................................................................717
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
...................................................................717
CALCulate<n>:MARKer<m>:X:SLIMits:ZOOM[:STATe]
..................................................... 718
..............................................................................................718
...................................................................................719
CALCulate<n>:MARKer<m>:LOEXclude
<State>
This command turns the suppression of the local oscillator during automatic marker
positioning on and off (for
all
markers in
all
windows).
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
CALC:MARK:LOEX ON
Manual operation:
See
CALCulate<n>:MARKer<m>:PEXCursion
<Excursion>
This command defines the peak excursion (for
all
markers in
all
windows).
The peak excursion sets the requirements for a peak to be detected during a peak
search.
The unit depends on the measurement.
Application/Result display
Unit
Spectrum
dB
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<Excursion>
The excursion is the distance to a trace maximum that must be
attained before a new maximum is recognized, or the distance to
a trace minimum that must be attained before a new minimum is
recognized
*RST:
6 dB in the Spectrum application and RF displays
Example:
CALC:MARK:PEXC 10dB
Defines peak excursion as 10 dB.