![Rohde & Schwarz FPL1000 Series User Manual Download Page 649](http://html.mh-extra.com/html/rohde-and-schwarz/fpl1000-series/fpl1000-series_user-manual_1477826649.webp)
Remote Commands
R&S
®
FPL1000
649
User Manual 1178.3370.02 ─ 03
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:STRack:BANDwidth
<Bandwidth>
This command defines the bandwidth around the center frequency that is included in
the signal tracking process.
Note that you have to turn on signal tracking before you can use the command.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<Bandwidth>
Range:
10 Hz to Max span
*RST:
(= span/10 on activating the function)
Default unit: Hz
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:STRack:THReshold
<Level>
This command defines the threshold level for the signal tracking process.
Note that you have to turn on signal tracking before you can use the command.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<Level>
The unit depends on
Range:
-130 dBm to 30 dBm
*RST:
-120 dBm
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:STRack:TRACe
<TraceNumber>
This command selects the trace on which the largest signal is searched for.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<TraceNumber>
1 to 6
Range:
1 to 6
*RST:
1
Setting Basic Spectrum RF Measurement Parameters