![Rohde & Schwarz FPL1000 Series User Manual Download Page 717](http://html.mh-extra.com/html/rohde-and-schwarz/fpl1000-series/fpl1000-series_user-manual_1477826717.webp)
Remote Commands
R&S
®
FPL1000
717
User Manual 1178.3370.02 ─ 03
Manual operation:
See
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
<State>
This command turns marker search limits on and off for
all
markers in
all
windows.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:MARK:X:SLIM ON
Switches on search limitation.
Manual operation:
See
" Search Limits ( Left / Right )"
See
" Deactivating All Search Limits "
See
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
<SearchLimit>
This command defines the left limit of the marker search range for
all
markers in
all
windows.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<SearchLimit>
The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST:
left diagram border
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
Manual operation:
See
" Search Limits ( Left / Right )"
See
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
<SearchLimit>
This command defines the right limit of the marker search range for
all
markers in
all
windows.
Setting Basic Spectrum RF Measurement Parameters