Remote Commands
R&S
®
FPL1000
596
User Manual 1178.3370.02 ─ 03
[SENSe:]LIST:RANGe<range>:LIMit:STATe
This command turns the limit check for all spurious emission measurement ranges on
and off (<range> is irrelevant).
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
[SENSe:]LIST:RANGe<range>:LIMit:STOP
<Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Level>
Absolute limit at the stop frequency of a SEM range.
Range:
-400 to 400
*RST:
13
Default unit: dBm
Manual operation:
See
" Abs Limit Start / Abs Limit Stop "
[SENSe:]LIST:RANGe<range>:POINts[:VALue]
<Points>
This command defines the number of sweep points in a spurious emission measure-
ment range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Points>
For more information on sweep points see
"How Much Data is Measured: Sweep Points and Sweep
Count"
*RST:
1001
Manual operation:
See
[SENSe:]LIST:RANGe<range>:RLEVel
<RefLevel>
This command defines the reference level for a spurious emission measurement
range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Configuring and Performing Measurements