Common Instrument Functions
R&S
®
FPL1000
99
User Manual 1178.3370.02 ─ 03
Access
: "Save" / "Open" icon in the toolbar
Or
: FILE
Possibly you would like to restore or repeat a measurement you performed under spe-
cific conditions on the instrument. Or you want to evaluate imported data in another
application on the R&S
FPL1000 and would like to restore the measurement settings
applied during measurement. In these cases, you can store and recall instrument and
measurement settings, and possibly other related measurement data.
Two different methods are available for managing instrument settings:
●
Quick Save/Quick Recall - a defined set of instrument settings or channel setups
are stored or recalled quickly in just one step
●
Configurable Save/Recall - a user-defined set of instrument settings or channel
setups are stored to a definable storage location
Restrictions when recalling measurement settings
When recalling a saved configuration file, the following restrictions apply:
●
The R&S
FPL1000 must support the frequency range defined in the configuration
file.
●
Configuration files created on a R&S
FPL1000 with certain options in use do not
work on an R&S
FPL1000 without these options.
●
Files created with newer firmware versions may not work with a previous version.
●
Files created on an instrument other than the R&S
FPL1000 do not work on the
R&S
FPL1000.
Data Management