Remote Commands
R&S
®
FPL1000
780
User Manual 1178.3370.02 ─ 03
Manual operation:
See
CALCulate<n>:LIMit<k>:DELete
This command deletes a limit line.
Suffix:
<n>
.
<k>
Usage:
Event
Manual operation:
See
CALCulate<n>:LIMit<k>:STATe
<State>
This command turns the limit check for a specific limit line on and off.
To query the limit check result, use
.
Note that a new command exists to activate the limit check and define the trace to be
checked in one step (see
CALCulate<n>:LIMit<k>:TRACe<t>:CHECk
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:LIM:STAT ON
Switches on the limit check for limit line 1.
Manual operation:
See
CALCulate<n>:LIMit<k>:TRACe<t>:CHECk
<State>
This command turns the limit check for a specific trace on and off.
To query the limit check result, use
.
Suffix:
<n>
.
<k>
<t>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0