The Spectrum Application (RF Measurements)
R&S
®
FPL1000
234
User Manual 1178.3370.02 ─ 03
Remote command:
All Functions Off
Switches off all measurement functions and returns to a basic frequency sweep.
Selecting "Frequency Sweep" has the same effect.
7.2.1.2
How to Perform a Basic Sweep Measurement
The following step-by-step instructions demonstrate how to perform basic sweep mea-
surements.
For remote operation, see
Chapter 8.6.14, "Programming Example: Performing a Basic
To perform one or more single sweeps
1. Configure the frequency and span to be measured ( "Frequency" dialog box, see
Chapter 7.4, "Frequency and Span Configuration"
2. Configure the number of sweeps to be performed in a single measurement
( "Sweep Config" dialog box, see
3. If necessary, configure a trigger for the measurement ( "Trigger/ Gate Config" dia-
log box, see
Chapter 7.7, "Trigger and Gate Configuration"
4. Define how the results are evaluated for display ( "Trace" dialog box, see
ter 7.10.1.2, "Trace Settings"
5. If necessary, configure the vertical axis of the display ( "Amplitude" dialog box, see
Chapter 7.5, "Amplitude and Vertical Axis Configuration"
6. To start the measurement, select one of the following:
● RUN SINGLE key
● "Single Sweep" softkey in the "Sweep" menu
The defined number of sweeps are performed, then the measurement is stopped.
While the measurement is running, the RUN SINGLE key is highlighted. To abort
the measurement, press the RUN SINGLE key again. The key is no longer high-
lighted. The results are not deleted until a new measurement is started.
7. To repeat the same number of sweeps without deleting the last trace, select the
"Continue Single Sweep" softkey in the "Sweep" menu.
Measurements and Results