Remote Commands
R&S
®
FPL1000
626
User Manual 1178.3370.02 ─ 03
CALCulate<n>:MARKer<m>:FUNCtion:MSUMmary
<TimeOffset>,<MeasTime>,<PulsePeriod>,<#OfPulses>
This command configures power measurements on pulses in the time domain.
To evaluate the pulse power, the R&S
FPL1000 uses the data captured during a previ-
ous measurement. The data recorded during the set measurement time is combined to
a measured value for each pulse according to the detector specified and the indicated
number of results is output as a list.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Parameters:
<TimeOffset>
Defines a time offset to start the measurement at the first pulse
of a trace.
*RST:
0
Default unit: s
<MeasTime>
Defines the measurement time.
Default unit: s
<PulsePeriod>
Defines the pulse period
<#OfPulses>
Defines the number of pulses to measure.
Example:
CALC:MARK:FUNC:MSUM 50US,450US,576.9US,8
Evaluates data that contains 8 pulses during a measurement
time of 450
µs and a pulse period of 576.9
µs. The evaluation
starts with an offset of 50
µs.
[SENSe:]MPOWer:FTYPe
<FilterType>
This command selects the filter type for pulse power measurements.
Configuring and Performing Measurements