Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:MARK:MAX:AUTO ON
Activates the automatic peak search function for marker 1 at the
end of each particular sweep.
Manual operation:
See
" Auto Max Peak Search / Auto Min Peak Search "
CALCulate<n>:MARKer<m>:MAXimum:LEFT
This command moves a marker to the next lower peak.
The search includes only measurement values to the left of the current marker posi-
tion.
In the spectrogram, the command moves a marker horizontally to the maximum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
<m>
Usage:
Event
Manual operation:
See
CALCulate<n>:MARKer<m>:MAXimum:NEXT
This command moves a marker to the next lower peak.
In the spectrogram, the command moves a marker horizontally to the maximum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
<m>
Usage:
Event
Manual operation:
See
CALCulate<n>:MARKer<m>:MAXimum[:PEAK]
This command moves a marker to the highest level.
In the spectrogram, the command moves a marker horizontally to the maximum level in
the currently selected frame. The vertical marker position remains the same.
If the marker is not yet active, the command first activates the marker.