![Rohde & Schwarz FPL1000 Series User Manual Download Page 20](http://html.mh-extra.com/html/rohde-and-schwarz/fpl1000-series/fpl1000-series_user-manual_1477826020.webp)
Getting Started
R&S
®
FPL1000
20
User Manual 1178.3370.02 ─ 03
Operating temperature
Before performing this functional test, make sure that the instrument has reached its
operating temperature (for details, refer to the data sheet).
A message in the status bar ( "Instrument warming up..." ) indicates that the operating
temperature has not yet been reached.
Depending on the installation settings, an automatic self-alignment is performed each
time the instrument is switched on, and a dialog is displayed indicating how much
warm-up time is still required before self-alignment can be performed.
Performing a self-alignment
1. Press the SETUP key.
2. Press the "Alignment" softkey.
3. Select the "Start Self Alignment" button in the "Alignment" dialog box.
Once the system correction values have been calculated successfully, a message
is displayed.
To display the alignment results again later
●
Press the SETUP key.
●
Press the "Alignment" softkey.
4.1.1.10
Checking the Supplied Options
The instrument may be equipped with both hardware and firmware options. To check
whether the installed options correspond to the options indicated on the delivery note,
proceed as follows.
1. Press the SETUP key.
2. Press the "System Config" softkey.
3. Switch to the "Ve Options" tab in the "System Configuration" dialog.
A list with hardware and firmware information is displayed.
4. Check the availability of the hardware options as indicated in the delivery note.
4.1.2 Windows Operating System
The instrument contains the Windows operating system which has been configured
according to the instrument's features and needs. Changes in the system setup are
only required when peripherals like keyboard or a printer are installed or if the network
configuration does not comply with the default settings. After the R&S
FPL1000 is star-
ted, the operating system boots and the instrument firmware is started automatically.
Preparing for Use