Getting Started
R&S
®
FPL1000
42
User Manual 1178.3370.02 ─ 03
The Ref. Out connectors can be used to provide a reference signal from the
R&S
FPL1000 to other devices that are connected to this instrument.
Various connectors are provided for different reference signals:
Connector
Reference signal
Usage
Ref. In
10 MHz
10 dBm
To provide an external reference signal on the
R&S
FPL1000.
Ref. Out
10 MHz
10 dBm
To provide the internal reference signal from the
R&S
FPL1000 to another device continuously.
For the R&S
FPL1000 this is also used to provide the
optional OCXO reference signal to another device.
4.2.2.5
Trigger In
Use the female Trigger In connector to input an external trigger or gate data. Thus, you
can control the measurement using an external signal. The voltage level is 1.4
V. The
typical input impedance is 10
kΩ.
4.2.2.6
DVI
You can connect an external monitor or other display device to the R&S
FPL1000 via
the DVI (Digital visual interface) connector to provide an enlarged display.
For details, see
Chapter 4.1.4, "Connecting an External Monitor"
4.2.2.7
LAN
The LAN interface can be used to connect the R&S
FPL1000 to a local network for
remote control, printouts or data transfer. The assignment of the RJ-45 connector sup-
ports twisted-pair category 5 UTP/STP cables in a star configuration (UTP stands for
unshielded twisted pair
, and STP for
shielded twisted pair
).
For details, see
Chapter 6.4, "Network and Remote Operation"
4.2.2.8
USB
The rear panel provides two additional female USB (3.0 standard) connectors to con-
nect devices like a keyboard, a mouse or a memory stick (see also
4.2.2.9
Sensor Connector
The LEMOSA female connector is used to connect power sensors of the R&S
NRP-
Zxy family. For a detailed list of supported sensors, see the data sheet.
For details on configuring and using power sensors, see
Instrument Tour