Common Instrument Functions
R&S
®
FPL1000
222
User Manual 1178.3370.02 ─ 03
How to Configure the R&S
FPL1000 for Remote Operation via Remote Desktop
1. Create a fixed IP address for the TCP/IP protocol as described in
Note:
To avoid problems, use a fixed IP address.
When a DHCP server is used, a new IP address is assigned each time the instru-
ment is restarted. This address must first be determined on the instrument itself.
Thus, using a DHCP server is not suitable for remote operation of the
R&S
FPL1000 via Remote Desktop.
2.
Select the "Windows" icon in the toolbar to access the operating system.
3. In the "Start" menu, select the "Control Panel" and then "System and Security".
4. In the "System" area, select "Allow remote access".
5. Define which users are to be given access to the R&S
FPL1000 via Remote Desk-
top.
Note
: The user account under which configuration is carried out is automatically
enabled for Remote Desktop.
a) In the "Remote" tab, in the "Remote Desktop" area, select the "Select Users"
button.
b) Select the users or create new user accounts as described in
c) Select "OK" to confirm the settings.
6. The R&S
FPL1000 is now ready for connection setup with the Remote Desktop
program of the controller.
Network and Remote Operation