This connector is only available if the R&S
FPL1-B5 option is installed.
For details, see
4.2.2.13
Noise Source Control
The Noise Source Control female connector is used to provide the supply voltage for
an external noise source. For example, use it to measure the noise figure and gain of
amplifiers and frequency converting devices.
This connector is only available if the R&S
FPL1-B5 option is installed.
Conventional noise sources require a voltage of +28
V to be switched on and 0
V to be
switched off. The output supports a maximum load of 100
mA.
4.3 Trying Out the Instrument
This chapter introduces the most important functions and settings of the R&S
FPL1000
step by step. The complete description of the functionality and its usage is given in the
R&S
FPL1000 User Manual. Basic instrument operation is described in
Prerequisites
●
The instrument is set up, connected to the mains system, and started up as descri-
bed in
Chapter 4.1, "Preparing for Use"
For these first measurements, you use the internal calibration signal, so you do not
need any additional signal source or instruments. Try out the following:
........................................................................................44
.......................................................................................46
Activating Additional Measurement Channels
.........................................................48
Performing Sequential Measurements
....................................................................52
.................................................................................. 53
.................................................................................55
.........................................................................................56
....................................................................................61
4.3.1 Measuring a Basic Signal
We will start out by measuring a simple sinus wave, using the internal calibration signal
as the input.
To display the internal 50 MHz calibration signal
1. Press the PRESET key to start out in a defined instrument configuration.
2. Press the SETUP key.
Trying Out the Instrument