Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:MODE
<Mode>
This command selects the type of scaling of the y-axis (for all traces).
When the display update during remote control is off, this command has no immediate
effect.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Mode>
ABSolute
absolute scaling of the y-axis
RELative
relative scaling of the y-axis
*RST:
ABSolute
Example:
DISP:TRAC:Y:MODE REL
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision
<Value>
This remote command determines the grid spacing on the Y-axis for all diagrams,
where possible.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Value>
numeric value WITHOUT UNIT (unit according to the result dis-
play)
Defines the range per division (total range = 10*<Value>)
*RST:
depends on the result display
Example:
DISP:TRAC:Y:PDIV 10
Sets the grid spacing to 10 units (e.g. dB) per division
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
<Position>
This command defines the vertical position of the reference level on the display grid
(for all traces).
The R&S
FPL1000 adjusts the scaling of the y-axis accordingly.
Suffix:
<n>
.
Setting Basic Spectrum RF Measurement Parameters