Remote Commands
R&S
®
FPL1000
600
User Manual 1178.3370.02 ─ 03
CALCulate<n>:PSEarch:SUBRanges
<NumberPeaks>
CALCulate<n>:PEAKsearch:SUBRanges
<NumberPeaks>
This command defines the number of peaks included in the peak list.
After this number of peaks has been found, the R&S
FPL1000 stops the peak search
and continues the search in the next measurement range.
Suffix:
<n>
.
Parameters:
<NumberPeaks>
Range:
1 to 50
*RST:
25
Example:
CALC:PSE:SUBR 10
Sets 10 peaks per range to be stored in the list.
Manual operation:
See
8.6.7.4
Performing a Spurious Measurement
The following commands are required to perform a Spurious measurement:
SENS:SWE:MODE LIST
, see
Chapter 8.6.1, "Performing Measure-
8.6.7.5
Retrieving and Saving Settings and Results
The following commands analyze and retrieve measurement results for Spurious mea-
surements.
Useful commands for spurious emission measurements described elsewhere
●
●
●
●
8.6.7.6
Programming Example: Spurious Emissions Measurement
In the following example, the Spurious Emissions measurement is configured by defin-
ing ranges and parameters to create the following sweep list.
Note that this example is primarily meant to demonstrate the remote control com-
mands, it does not necessarily reflect a useful measurement task.
//------------Preparing the measurement-------------
*RST
//Resets the instrument
Configuring and Performing Measurements