Remote Commands
R&S
®
FPL1000
536
User Manual 1178.3370.02 ─ 03
Example:
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
INIT:CONM;*WAI
Continues the measurement (next 20 sweeps) and waits for the
end.
Result: Averaging is performed over 40 sweeps.
Usage:
Event
Manual operation:
See
INITiate<n>:CONTinuous
<State>
This command controls the sweep mode for an individual channel setup.
Note that in single sweep mode, you can synchronize to the end of the measurement
with *OPC, *OPC? or *WAI. In continuous sweep mode, synchronization to the end of
the measurement is not possible. Thus, it is not recommended that you use continuous
sweep mode in remote control, as results like trace data or markers are only valid after
a single sweep end synchronization.
Example:
INIT:CONT OFF
Switches the sweep mode to single sweep.
INIT:CONT ON
Switches the sweep mode to continuous sweep.
Manual operation:
See
See
See
" Continuous Sweep / Run Cont "
INITiate<n>[:IMMediate]
This command starts a (single) new measurement.
With sweep count or average count > 0, this means a restart of the corresponding
number of measurements. With trace mode MAXHold, MINHold and AVERage, the
previous results are reset on restarting the measurement.
You can synchronize to the end of the measurement with *OPC, *OPC? or *WAI.
Suffix:
<n>
.
irrelevant
Configuring and Performing Measurements