![Rohde & Schwarz FPL1000 Series User Manual Download Page 546](http://html.mh-extra.com/html/rohde-and-schwarz/fpl1000-series/fpl1000-series_user-manual_1477826546.webp)
Remote Commands
R&S
®
FPL1000
546
User Manual 1178.3370.02 ─ 03
Manual operation:
See
[SENSe:]POWer:ACHannel:NAME:CHANnel<ch>
<Name>
This command defines a name for a transmission channel.
Suffix:
<ch>
.
1 to 18
Tx channel number
Parameters:
<Name>
String containing the name of the channel
*RST:
TX<1...12>
Manual operation:
See
[SENSe:]POWer:ACHannel:SPACing[:ACHannel]
<Spacing>
This command defines the distance from transmission channel to adjacent channel.
Parameters:
<Spacing>
Range:
100 Hz to 2000 MHz
*RST:
14 kHz
Manual operation:
See
[SENSe:]POWer:ACHannel:SPACing:ALTernate<ch>
<Spacing>
This command defines the distance from transmission channel to alternate channels.
If you set the channel spacing for the first alternate channel, the R&S
FPL1000 adjusts
the spacing of alternate channels of a lower order, but not the other way round. The
command works hierarchically: to set a distance from the transmission channel to the
2nd and 3rd alternate channel, you have to define a spacing for the 2nd alternate
channel first.
Suffix:
<ch>
.
1 to 11
Alternate channel number
Parameters:
<Spacing>
Range:
100 Hz to 2000 MHz
*RST:
40 kHz (ALT1), 60 kHz (ALT2), 80 kHz (ALT3), ...
Manual operation:
See
[SENSe:]POWer:ACHannel:SPACing:CHANnel<ch>
<Spacing>
This command defines the distance between transmission channels.
Configuring and Performing Measurements