Getting Started
R&S
®
FPL1000
50
User Manual 1178.3370.02 ─ 03
In the "Frequency" dialog box, set the
center frequency
to
50
MHz
and tap "Zero
Span" .
6. Increase the reference level so you can see the level of the calibration signal at
1
dBm.
a) In the "Amplitude" dialog box, set the "Reference Level" value to
+10
dBm
.
As the calibration signal does not vary over time, the level versus time diagram dis-
plays a straight line.
Figure 4-10: Time domain display of the calibration signal
7. Create a new measurement channel for I/Q analysis:
a) Press the MODE key.
b) Tap the "IQ Analyzer" button to activate a measurement channel for the I/Q
Analyzer application.
c) Tap the "Display Config" softkey to activate the SmartGrid mode.
Trying Out the Instrument