Remote Commands
R&S
®
FPL1000
660
User Manual 1178.3370.02 ─ 03
INPut:GAIN:STATe
<State>
This command turns the internal preamplifier on and off. It requires the optional pream-
plifier hardware.
The input signal is amplified by 20
dB if the preamplifier option is activated.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
INP:GAIN:STAT ON
Switches on 20 dB preamplification.
Manual operation:
See
8.8.3.4
Scaling the Y-Axis
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
................................................................... 660
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
............................................... 660
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:MODE
.........................................................661
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision
.................................................... 661
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
................................................... 661
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing
.................................................................. 662
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
<Range>
This command defines the display range of the y-axis (for all traces).
Note that the command works only for a logarithmic scaling. You can select the scaling
with
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing
.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Range>
Range:
1 dB to 200 dB
*RST:
100 dB
Example:
DISP:TRAC:Y 110dB
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
Automatic scaling of the y-axis is performed once, then switched off again (for all
traces).
Suffix:
<n>
.
<t>
irrelevant
Setting Basic Spectrum RF Measurement Parameters