![Rohde & Schwarz FPL1000 Series User Manual Download Page 287](http://html.mh-extra.com/html/rohde-and-schwarz/fpl1000-series/fpl1000-series_user-manual_1477826287.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
287
User Manual 1178.3370.02 ─ 03
8. Continue in this way until all carriers have been measured.
7.2.5.5
Measurement Example
In the following example, the bandwidth that occupies 99
% of the total power of a PDC
signal at 800
MHz, level 0
dBm is measured.
A programming example demonstrating an OBW measurement in a remote environ-
ment is provided in
Chapter 8.6.6.10, "Example: SEM Measurement"
1. Preset the R&S
FPL1000.
2. Set the "Center Frequency" to
800
MHz
.
3. Set the "Reference Level" to
-10
dBm
.
4. Press the MEAS key or select "Select Measurement" in the "Overview" .
5. Select the "OBW" measurement function.
6. Select the "OBW Config" softkey.
7. Set the "% Power Bandwidth" to
99
%
.
8. Set the "Channel Bandwidth" to
21
kHz
as specified by the PDC standard.
9. Optimize the settings for the selected channel configuration by selecting "Adjust
Settings" .
10. Adjust the reference level to the measured total power by selecting "Auto Level" in
the AUTO SET menu.
11. The PDC standard requires the peak detector for OBW measurement. In the
"Traces" configuration dialog, set the trace detector to "PositivePeak" .
12. Start a sweep.
The result is displayed as OBW in the marker results.
7.2.6 Spectrum Emission Mask (SEM) Measurement
Spectrum Emission Mask (SEM) measurements monitor compliance with a spectral
mask.
........................................................................................288
.............................................................................................. 288
................................................................................................301
How to Perform a Spectrum Emission Mask Measurement
..................................315
Measurement Example: Multi-SEM Measurement
................................................319
Reference: SEM File Descriptions
........................................................................ 320
Measurements and Results