10 Troubleshooting
If the results do not meet your expectations, the following sections may contain helpful
hints and information.
Error Messages in Remote Control Mode
.............................................................858
Troubleshooting Remote Operation
......................................................................859
Miscellaneous Troubleshooting Hints
................................................................... 860
Collecting Information for Technical Support
........................................................ 862
10.1 Error Information
If errors or irregularities are detected, a keyword and an error message, if available,
are displayed in the status bar.
Depending on the type of message, the status message is indicated in varying colors.
Table 10-1: Status bar information - color coding
Color
Type
Description
Red
Error
An error occurred at the start or during a measurement, e.g. due to missing
data or wrong settings, so that the measurement cannot be started or com-
pleted correctly.
Orange
Warning
An irregular situation occurred during measurement, e.g. the settings no lon-
ger match the displayed results, or the connection to an external device was
interrupted temporarily.
Gray
Information
Information on the status of individual processing steps.
No color
No errors
No message displayed - normal operation.
Green
Measurement
successful
Some applications visualize that the measurement was successful by show-
ing a message.
If any error information is available for a channel setup, an exclamation mark is dis-
played next to the channel setup name ( ). This is particularly useful when the Multi-
View tab is displayed, as the status bar in the MultiView tab always displays the infor-
mation for the currently selected measurement only.
Table 10-2: List of keywords
IF OVLD
Overload of the IF signal path in the A/D converter or in the digital IF.
●
Increase the reference level.
INPUT OVLD
The signal level at the RF input connector exceeds the maximum.
The RF input is disconnected from the input mixer to protect the device. To re-enable
measurement, decrease the level at the RF input connector and reconnect the RF
input to the mixer input.
Error Information