Remote Commands
R&S
®
FPL1000
564
User Manual 1178.3370.02 ─ 03
INITiate<n>:ESPectrum
This command initiates a Spectrum Emission Mask measurement.
Suffix:
<n>
.
irrelevant
Usage:
Event
[SENSe:]SWEep:MODE
<Mode>
This command selects the spurious emission and spectrum emission mask measure-
ments.
You can select other measurements with
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe]
Parameters:
<Mode>
AUTO
Turns on basic spectrum measurements.
ESPectrum
Turns on spectrum emission mask measurements.
LIST
Turns on spurious emission measurements.
*RST:
AUTO
8.6.6.3
Configuring a Multi-SEM Measurement
In the Spectrum application only, spectrum emissions can be measured for multiple
sub blocks of channels (see
"SEM with Multiple Sub Blocks ("Multi-SEM")"
on page 297). Up to three sub blocks (with two gaps) can be defined. For each sub
block, the familiar configuration settings concerning ranges, limit lines etc. can be
defined in individual tabs (select the sub block using the <sb> suffix in the correspond-
ing commands). In addition, settings on the sub blocks themselves must be configured.
Useful commands for multi-SEM measurements described elsewhere:
●
[SENSe:]ESPectrum<sb>:RANGe<range>:MLCalc
Remote commands exclusive to multi-SEM measurements
[SENSe:]ESPectrum<sb>:SBCenter
................................................................................ 564
..................................................................................564
................................................................................. 565
...................................................................................565
[SENSe:]ESPectrum<sb>:SBCenter
<Frequency>
[SENSe:]ESPectrum<sb>:SCENter
<Frequency>
This command defines the center frequency of the selected sub block in a Multi-SEM
measurement.
Configuring and Performing Measurements