Remote Commands
R&S
®
FPL1000
768
User Manual 1178.3370.02 ─ 03
Example:
CALC:DLIN2:STAT ON
Turns on display line 2.
CALCulate<n>:FLINe<k>
<Frequency>
This command defines the position of a frequency line.
Suffix:
<n>
.
<k>
Parameters:
<Frequency>
Note that you can not set a frequency line to a position that is
outside the current span.
Range:
0 Hz to Fmax
*RST:
(STATe to OFF)
Example:
CALC:FLIN2 120MHz
Sets frequency line 2 to a frequency of 120 MHz.
Manual operation:
See
" Vertical Line 1 / Vertical Line 2 "
CALCulate<n>:FLINe<k>:STATe
<State>
This command turns a frequency line on and off
Suffix:
<n>
.
<k>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:FLIN2:STAT ON
Turns frequency line 2 on.
CALCulate<n>:TLINe<k>
<Time>
This command defines the position of a time line.
Suffix:
<n>
.
<k>
Parameters:
<Time>
Note that you can not set a time line to a position that is higher
than the current sweep time.
Range:
0 s to 1600 s
*RST:
(STATe to OFF)
Setting Basic Spectrum RF Measurement Parameters