The Spectrum Application (RF Measurements)
R&S
®
FPL1000
335
User Manual 1178.3370.02 ─ 03
Remote command:
[SENSe:]LIST:RANGe<range>:LIMit:STATe
Abs Limit Start / Abs Limit Stop
Sets an absolute limit value at the start or stop frequency of the range [dBm].
Remote command:
[SENSe:]LIST:RANGe<range>:LIMit:STARt
[SENSe:]LIST:RANGe<range>:LIMit:STOP
Insert before Range / Insert after Range
Inserts a new range to the left of the currently focused range (before) or to the right
(after). The range numbers of the currently focused range and all higher ranges are
increased accordingly. The maximum number of ranges is 30.
Delete Range
Deletes the currently focused range. The range numbers are updated accordingly.
Adjusting the X-Axis to the Range Definitions
Access
: "Overview" > "Select Measurement" > "Spurious Emissions" > "Adjust X-Axis"
The frequency axis of the measurement diagram can be adjusted automatically so that
the span of all sweep list ranges corresponds to the displayed span. Thus, the x-axis
range is set from the start frequency of the first sweep range to the stop frequency of
the last sweep range.
List Evaluation
Access
: "Overview" > "Select Measurement" > "Spurious Emissions" > "List Evalua-
tion"
Configure the contents and display of the result list.
Measurements and Results