The Spectrum Application (RF Measurements)
R&S
®
FPL1000
342
User Manual 1178.3370.02 ─ 03
7.2.8.4
Time Domain Power Configuration
Access
: "Overview" > "Select Measurement" > "Time Domain Power" > "Time Dom
Power Config"
The remote commands required to perform these tasks are described in
"Measuring the Time Domain Power"
.................................................................................................342
Results
Activates the power results to be evaluated from the displayed trace or a limited area
of the trace.
"Peak"
Peak power over several measurements (uses trace averaging, Max
Hold)
"RMS"
RMS value from the points of the displayed trace or a segment
thereof.
"Mean"
Mean value from the points of the displayed trace or a segment
thereof. The linear mean value of the equivalent voltages is calcula-
ted.
"Std Dev"
The standard deviation of the measurement points from the mean
value.
The measurement of the mean power is automatically switched on at
the same time.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak[:STATe]
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:RESult?
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS[:STATe]
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:RESult?
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN[:STATe]
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:RESult?
Limit State
Switches the limitation of the evaluation range on or off. Default setting is off.
If deactivated, the entire sweep time is evaluated. If switched on, the evaluation range
is defined by the left and right limit. If only one limit is set, it corresponds to the left limit
and the right limit is defined by the stop frequency. If the second limit is also set, it
defines the right limit.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
Left Limit / Right Limit
Defines a power level limit for line S1 (left) or S2 (right).
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
Measurements and Results