Remote Commands
R&S
®
FPL1000
583
User Manual 1178.3370.02 ─ 03
<sb>
1 | 2 | 3
Parameters:
<NoPowerClasses> 1 to 4
*RST:
1
Example:
CALC:LIM:ESP:PCL:COUN 2
Two power classes can be defined.
Manual operation:
See
" Adding or Removing a Power Class "
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass<PClass>[:EXCLusive]
<State>
This command selects the power class used by the measurement if
is set to manual.
Note that:
●
You can only use power classes for which limits are defined.
Suffix:
<n>, <k>
.
irrelevant
<sb>
1 | 2 | 3
<PClass>
1...4
power class
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:LIM:ESP:PCL1 ON
Activates the first defined power class.
Manual operation:
See
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass<PClass>:LIMit[:STATe]
<State>
This command selects the limit check mode for each power class.
Suffix:
<n>, <k>
.
irrelevant
<sb>
1 | 2 | 3
<PClass>
1...4
power class
Configuring and Performing Measurements