Chapter 18
Troubleshooting
250
Other
Information
Symptom
Possible causes
Action
Connect the external monitor.
Images are not displayed on
the embedded monitor. (when
the external monitor is not
connected)
Connection to the external monitor is
removed in a state where images are
output to only the external monitor.
Reset the output setting of the embedded monitor and
external monitor. (Turn on the power of the DS-L3
again, turn off it while the Power indicator is blinking,
and then turn on the power again.)
Cables are connected incorrectly.
Use the commercial DVI cable (digital/analog) to
connect the DVI-I connector of the DS-L3 to the video
input connector of the external monitor.
Images are not displayed on
the external monitor.
The external monitor does not support
the video format of the DS-L3.
Use an XGA or SXGA-capable external monitor.
A noise shows up in the edge of
the screen when projecting
images onto a monitor or
projector.
The phase of output signals does not
accord with the display device.
Press the [LCD CHART] button in the [SETUP
MENU : ADDITIONAL] window to display LCD charts
for a monitor, and adjust a clock phase and clock
pitch in the display device (for details on the
adjustment, see the instructions manual of the display
device). To exit, press the [X] button on the top right.
18.3
Menu Operation
Symptom
Possible causes
Action
The embedded monitor is operated
without using a stylus.
Use the attached stylus for operating the embedded
monitor. When the stylus cannot be prepared,
operate the screen by a light touch with your
fingertips. Do not touch the screen with a sharp or
hard object.
The screen cannot be
operated.
The display of the external monitor
cannot be operated.
When images are output to only the external monitor,
connect a USB mouse to the DS-L3 and then operate
the display.
18.4
View of Images
Symptom
Possible cause
Action
The optical device is out of focus.
Bring the optical device into focus by operating the
focus drive or focusing ring.
The aperture diaphragm is closed too
much when using the microscope.
Adjust the aperture diaphragm to 70 to 80 percent of
the numerical aperture of the objective, or 70 to 80
percent of the pupil of the objective.
The magnifying power of the optical
device is too much.
As a guide, use 500 to 1000 times the numerical
aperture of the objective for the magnifying power.
The electronic zoom is being used for
observation.
Use of the electronic zoom may cause an image to
blur depending on the magnification.
Revise the setting of [SHARPNESS] in the [CAMERA
MENU : IMAGE] window.
The sharpness setting is low.
For the external monitor, adjust the sharpness setting
for the monitor.
The system is used in a place subject
to frequent vibration.
Use the system in a stabile, vibration-free place.
The image blurs.
The exposure time is too long.
If setting the exposure time long for a subject in
motion, the subject is blurred. Change the exposure
mode to shutter priority or manual, and set the
exposure time to a fast value.
When there is difficulty in focusing a dark subject, it is
recommended to use focus priority.
Summary of Contents for DS-L3
Page 2: ......