10 - 13
Note: N: normal beat; V: ventricular beat
10.7.3
Displaying Arrhythmia Information
You can display the arrhythmia information in the numeric area. To do so, follow this procedure:
1.
Access
Tile Layout
by either of the following ways:
◆
Select the
Screen Setup
quick key
→
select the
Tile Layout
tab.
◆
Select
Main Menu
quick key
→
from the
Display
column select
Tile Layout
.
2.
Click the numeric area where you want to display the arrhythmia information, and then select ECG
→
Arrhythmia
.
10.7.4
Changing Arrhythmia Settings
10.7.4.1
Changing Arrhythmia Alarm Settings
To set the arrhythmia alarm properties, follow this procedure:
1.
Select the ECG numeric area or waveform area to enter the
ECG
menu.
2.
Select the
Arrhythmia
tab
→
Alarm
tab.
3.
Enter the password if required. For more information, refer to
27.12 The Authorization Setup Settings
.
4.
Set alarm properties as desired.
NOTE
•
You can switch off lethal arrhythmia alarms only when you have enabled Lethal Arrhys Off. For more
information, see
10.7.4.2 Setting the Lethal Arrhythmia Alarms Switch
.
•
The priority of lethal arrhythmia alarms is always high. It cannot be altered.
10.7.4.2
Setting the Lethal Arrhythmia Alarms Switch
You can choose whether switching off lethal arrhythmia alarms is permissible or not. This function is password
protected. For more information, see
WARNING
•
If you switch off all arrhythmia alarms, the monitor will not alarm for any arrhythmia event. This may
result in a hazard to the patient. Always keep the patient under close surveillance.
NOTE
•
If any of the lethal arrhythmia alarms is switched off, the ECG waveform area displays the “Lethals
Off” message.
Pause
No QRS complex is detected within the set time threshold of pause.
Irr Rhythm
Consistently irregular rhythm (N, irregular RR interval change is greater than 12.5%)
A-Fib
P wave is absent and normal beat RR intervals are irregular.
PVCs/min
PVCs/min exceeds high limit.
Pauses/min
Pauses/min exceeds high limit.
Irr Rhythm End
Irregular rhythm no longer detected for the irregular rhythm end delay time.
Afib End
Atrial fibrillation no longer detected for the Afib end delay time.
Arrhythmia message
Description
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...